Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
CISA SSVCTrack at low or medium mission impact; Track or Attend at high (mission-essential systems).
No fix is confirmed yet. Restrict network exposure of the affected system or apply the vendor's mitigation within your standard update timelines at low or medium mission impact and sooner than that at high, and watch the vendor's advisory for the fix.
Exploitation none (no KEV listing, exploit record or EPSS ≥ 50%) · Automatable unknown (not published for this CVE) · Technical impact partial (CVSS below 9.0). Mission impact is CISA's Mission & Well-being decision point, and only you can judge it: high means the affected system is essential to your organisation's mission, or its compromise could cause irreversible harm to people. CISA's decision table